Operation Summary

This manual covers the operation of the Panasonic Ultra-Low Temperature Freezer, model MDF-C8V1 Series. It details installation procedures, start-up, temperature setting, and key lock functions. Information on alarm settings, compressor delay time, and routine maintenance like cleaning and defrosting is also provided. Specific sections address optional temperature recorders and backup cooling kits. Troubleshooting guidance and performance specifications are included to ensure proper and safe usage of the unit.

Troubleshooting and error codes

Situation Indication Buzzer Safety operation
High temperature alarm If the chamber temperature is higher than the temperature at which the high temperature alarm is activated. Intermittent tone with 15 minutes delay. Remote alarm with 15 minutes delay.
Low temperature alarm If the chamber temperature is lower than the temperature at which the low temperature alarm is activated. Intermittent tone with 15 minutes delay. Remote alarm with 15 minutes delay.
Power failure alarm In the case of power failure. If the power switch is OFF. When the power to the unit is disconnected. Intermittent tone Remote alarm.
Sensor abnormality If the thermal sensor is disconnected.. If the thermal sensor is short-circuited If the thermal sensor of the compressor is disconnected. If the thermal sensor of the compressor is short-circuited Alarm lamp is flashed. E01 and chamber temp. are displayed alternately. Alarm lamp is flashed. E02 and chamber temp. are displayed alternately. Alarm lamp is flashed. E05 and chamber temp. are displayed alternately. Alarm lamp is flashed. E06 and chamber temp. are displayed alternately. Intermittent tone
Remote alarm.
Compressor temp. abnormality In the event of failure of fan motor for cooling the compressor. Alarm lamp is flashed. E10 and chamber temp. are displayed alternately. Intermittent tone
Compressor stops.
Fan motor abnormality In the event of failure of fan motor for cooling the compressor. Alarm lamp is flashed. E15 and chamber temp. are displayed alternately. Intermittent tone
Remote alarm.
Battery switch check When battery switch is OFF at the time of alarm test. Alarm lamp is flashed. E09 is flashed. ----
Battery check When approx. about 3 years has passed with the power cord connected to the outlet. Battery check lamp lights. ----
Fan motor check When about 6 years has passed with power switch ON. Battery check lamp is flashed. Intermittent tone
Remote alarm.