Starting Up the Unit:
Operation After Power Failure:
The freezer resumes operation with the previous settings. A high-temperature alarm will activate if the chamber temperature is higher than the preset when power is restored.
| Alarm & Safety | Situation | Indication | Buzzer | Safety operation |
|---|---|---|---|---|
| High temperature alarm | If the chamber temperature is higher than the temperature at which the high temperature alarm is activated. | Alarm lamp is flashed. Digital temperature indicator is flashed. | Intermittent tone with about 15 minutes delay. | Remote alarm with about 15 minutes delay. |
| Low temperature alarm | If the chamber temperature is lower than the temperature at which the low temperature alarm is activated. | Alarm lamp is flashed. | Intermittent tone | Remote alarm. |
| Power failure alarm | When the power to the unit is disconnected. | Alarm lamp is flashed. | Intermittent tone | Remote alarm. |
| Filter alarm | When the condenser filter is clogged. | Filter check lamp lights. | Intermittent tone | Remote alarm. |
| Battery check | When about 3 years has passed with power switch ON. | Battery check lamp lights. | ----- | Unit keeps continuous running. |
| Fan motor check | When about 6 years has passed with power switch ON. | Battery check lamp is flashed. Chamber temperature is displayed. | ----- | Remote alarm. Unit keeps continuous running. |
| Auto-return | When there is no key pressing in each setting mode for 90 seconds. | Chamber temperature is displayed. | ----- | Finishing of each setting mode. |
| Key lock | When the key lock is ON. | Alarm lamp is flashed. E01 and chamber temp. are displayed alternately. | Intermittent tone | Change of setting is disable. Remote alarm. |
| Sensor abnormality | If the thermal sensor is disconnected. | Alarm lamp is flashed. E01 and chamber temp. are displayed alternately. | Intermittent tone | Remote alarm. |
| If the thermal sensor is short-circuited. | Alarm lamp is flashed. E02 and chamber temp. are displayed alternately. | Intermittent tone | Remote alarm. | |
| If the cascade sensor is disconnected. | Alarm lamp is flashed. E03 and chamber temp. are displayed alternately. | Intermittent tone | Remote alarm. | |
| If the cascade sensor is short circuited. | Alarm lamp is flashed. E04 and chamber temp. are displayed alternately. | Intermittent tone | Remote alarm. | |
| If the filter sensor is disconnected. | Alarm lamp is flashed. E05 and chamber temp. are displayed alternately. | Intermittent tone | Remote alarm. | |
| If the filter sensor is short-circuited. | Alarm lamp is flashed. E06 and chamber temp. are displayed alternately. | Intermittent tone | Remote alarm. | |
| If the ambient temperature sensor is disconnected. | Alarm lamp is flashed. E07 and chamber temp. are displayed alternately. | Intermittent tone | Remote alarm. | |
| Battery switch check | When the battery switch is OFF during alarm test. | Alarm lamp is flashed. E09 is flashed. | ----- | ----- |
| If the ambient temperature sensor is short-circuited. | Alarm lamp is flashed. E08 and chamber temp. are displayed alternately. | Intermittent tone | Remote alarm. | |
| Condenser temp. abnormality | When the fan motor has broken down | E10 and chamber temp. are displayed alternately. | Intermittent tone | Compressor of high stage side stops. |