Starting Up the Unit:

  1. Turn off the backup cooling kit (if installed).
  2. Connect the power cord to a dedicated outlet and turn on the freezer's power switch.
  3. Turn on the battery switch.
  4. Set the desired chamber temperature.
  5. Verify the chamber temperature reaches the set point.
  6. Turn on the backup cooling kit (if installed).
  7. Test the alarm system by pressing the ALARM TEST key.
  8. Once confirmed, load items into the chamber in small batches.

Operation After Power Failure:

The freezer resumes operation with the previous settings. A high-temperature alarm will activate if the chamber temperature is higher than the preset when power is restored.

Troubleshooting and error codes

Alarm & SafetySituationIndicationBuzzerSafety operation
High temperature alarmIf the chamber temperature is higher than the temperature at which the high temperature alarm is activated.Alarm lamp is flashed. Digital temperature indicator is flashed.Intermittent tone with about 15 minutes delay.Remote alarm with about 15 minutes delay.
Low temperature alarmIf the chamber temperature is lower than the temperature at which the low temperature alarm is activated.Alarm lamp is flashed.Intermittent toneRemote alarm.
Power failure alarmWhen the power to the unit is disconnected.Alarm lamp is flashed.Intermittent toneRemote alarm.
Filter alarmWhen the condenser filter is clogged.Filter check lamp lights.Intermittent toneRemote alarm.
Battery checkWhen about 3 years has passed with power switch ON.Battery check lamp lights.-----Unit keeps continuous running.
Fan motor checkWhen about 6 years has passed with power switch ON.Battery check lamp is flashed.
Chamber temperature is displayed.
-----Remote alarm.
Unit keeps continuous running.
Auto-returnWhen there is no key pressing in each setting mode for 90 seconds.Chamber temperature is displayed.-----Finishing of each setting mode.
Key lockWhen the key lock is ON.Alarm lamp is flashed.
E01 and chamber temp. are displayed alternately.
Intermittent toneChange of setting is disable.
Remote alarm.
Sensor abnormalityIf the thermal sensor is disconnected.Alarm lamp is flashed.
E01 and chamber temp. are displayed alternately.
Intermittent toneRemote alarm.
If the thermal sensor is short-circuited.Alarm lamp is flashed.
E02 and chamber temp. are displayed alternately.
Intermittent toneRemote alarm.
If the cascade sensor is disconnected.Alarm lamp is flashed.
E03 and chamber temp. are displayed alternately.
Intermittent toneRemote alarm.
If the cascade sensor is short circuited.Alarm lamp is flashed.
E04 and chamber temp. are displayed alternately.
Intermittent toneRemote alarm.
If the filter sensor is disconnected.Alarm lamp is flashed.
E05 and chamber temp. are displayed alternately.
Intermittent toneRemote alarm.
If the filter sensor is short-circuited.Alarm lamp is flashed.
E06 and chamber temp. are displayed alternately.
Intermittent toneRemote alarm.
If the ambient temperature sensor is disconnected.Alarm lamp is flashed.
E07 and chamber temp. are displayed alternately.
Intermittent toneRemote alarm.
Battery switch checkWhen the battery switch is OFF during alarm test.Alarm lamp is flashed.
E09 is flashed.
----------
If the ambient temperature sensor is short-circuited.Alarm lamp is flashed.
E08 and chamber temp. are displayed alternately.
Intermittent toneRemote alarm.
Condenser temp. abnormalityWhen the fan motor has broken downE10 and chamber temp. are displayed alternately.Intermittent toneCompressor of high stage side stops.