  • Turn on the power switch.
  • Press the mode change key to enter temperature set mode.
  • Use the numerical value shift keys to adjust the digits of the desired temperature.
  • Use the digit shift key to move between digits.
  • Press the enter key to confirm the new temperature setting.
  • The unit automatically returns to display mode after 90 seconds of inactivity.

Troubleshooting and error codes

Malfunction Possible Cause Solution
No refrigeration The voltage is too low. Call an electrician.
The breaker is free. Check the breaker.
The fuse is blown. Replace the fuse.
The freezer is not cooled at all for about 5 minutes after starting-up. This is normal.
Poor refrigeration The ambient temperature is too high. Ensure adequate ventilation.
The door is not shut tightly. Check door seal and close properly.
The inner lid is not installed correctly. Reinstall the inner lid.
The set temperature in the controller is not set properly. Adjust the set temperature.
The grille is blocked out. Clean the grille.
The filter is clogged. Clean the filter when the filter check lamp lights.
The freezer is in the direct sunlight. Move the freezer to a shaded location.
There is any heating source near the freezer. Move the freezer away from heat sources.
A rubber cap and insulation for the access port are not set correctly. Reinstall the rubber cap and insulation.
You put too many unfrozen articles into the freezer compartment. Load items gradually.
Too much frost is accumulated in the chamber. Defrost the freezer.
Noise The freezer is not installed on the sturdy floor. Install on a sturdy floor.
There is anything touching the frame. Ensure nothing is touching the frame.
The freezer is not leveled with the leveling feet. Adjust the leveling feet.
It is the immediately after starting of refrigeration. This is normal.
The freezer causes larger noise when the chamber temperature is high because of large load. Noise gets smaller along with chamber temperature.
Alarm test key cannot actuate the alarm function The alarm test key does not function when the power switch of the freezer is OFF. Ensure the power switch is ON.
When only the buzzer or only the alarm is actuated by the alarm test key, the unactuated part is out of order, and must be replaced. Contact service.
Recorder does not operate normally (MDF-192AT only) If the chart is not advanced, examine the dry cell. Replace the cell for recorder every 1 year.
If the temperature is not recorded, examine the recording pen. The pen should be our specified one.
Back-up test switch does not operate normally (MDF-192AT only) The liquid CO2 cylinder is empty. Replace the cylinder.
The valve of the liquid CO2 cylinder is not opened. Open the valve.
The ambient temperature is too high. Move the liquid CO2 cylinder to a cool location. Inquire at liquid CO2 suppliers about cylinder check, adjustment, installation, or move.