Message | Purpose of Message | Type | Action Required |
---|---|---|---|
# of Tests has been set to min | The operator has attempted to set the number of Online Clearance (OLC) tests lower than allowed. | Dialog Message | The machine has set the number of tests to the lowest value allowed. Verify that the number of OLC tests is acceptable. See page 144 for more information. |
5 Minutes Flow Off | Dialysate flow has been off for five minutes. | Low Alarm | Press Reset to silence the alarm. If no dialysate flow is desired, continue without running dialysate flow, see “Isolated Ultrafiltration” on page 99 for more information. If you intend for the flow to be on, set the Dialysate Flow button to the prescribed rate and press the CONFIRM key to resume dialysate flow. |
**** 5V High **** | Electronic self-test, power supply limits exceeded. | High Alarm | Turn machine power off and back on. If alarm is not cleared, take the machine out of service and alert a qualified service technician. |
Message | Purpose of Message | Type | Action Required |
---|---|---|---|
# of Tests has been set to min | The operator has attempted to set the number of Online Clearance (OLC) tests lower than allowed. | Dialog Message | The machine has set the number of tests to the lowest value allowed. Verify that the number of OLC tests is acceptable. See page 144 for more information. |
5 Minutes Flow Off | Dialysate flow has been off for five minutes. | Low Alarm | Press Reset to silence the alarm. If no dialysate flow is desired, continue without running dialysate flow, see “Isolated Ultrafiltration” on page 99 for more information. If you intend for the flow to be on, set the Dialysate Flow button to the prescribed rate and press the CONFIRM key to resume dialysate flow. |
**** 5V High **** | Electronic self-test, power supply limits exceeded. | High Alarm | Turn machine power off and back on. If alarm is not cleared, take the machine out of service and alert a qualified service technician. |