Warranty Summary
Message | What it means | What to do |
---|---|---|
Er-1 |
A used test strip was inserted. | Remove the test strip and insert a new test strip. |
Er-2 |
The blood or control solution sample was applied before the symbol appeared. | Repeat the test with a new test strip and wait until the symbol appears before applying the blood or control solution sample. |
Er-3 |
The thermometer of meter is not stabilized yet. | Retry the test after about 30 minutes for the thermometer to be stabilized. |
Er-4 |
The blood sample is of insufficient volume or blood is too thick. | Repeat the test using a new test strip. |
Er-5 |
This error message may appear when the wrong blood glucose test strip is used instead of the ReliOn™ Premier Test Strip. | Repeat the test with a ReliOn™ Premier Test Strip. Insert the test strip with the contact bars facing up, and push in gently until the meter turns on. |
Er-6 |
There is a problem with the meter. | Do not use the meter. Contact Customer Service: 1-855-776-0662. |
Er-7 |
There is a problem with the meter. This error message may also appear when the wrong blood glucose test strip is used instead of the ReliOn™ Premier Test Strip. | Repeat the test with a ReliOn™ Premier Test Strip. If the correct test strip is inserted, the meter will automatically turn on and enter test mode. If the error message persists even after inserting the correct test strip, do not use the meter and contact Customer Service: 1-855-776-0662. |
Er-8 |
There is a problem with the meter. | Do not use the meter. Contact Customer Service: 1-855-776-0662. |
Er-9 |
There is a problem with the meter. | Do not use the meter. Contact Customer Service: 1-855-776-0662. |
LoF |
The temperature during the test is below the operating range. | Move to an area where the temperature is within the operating range (50–104 ˚F) and repeat the test after the meter and test strips have reached a temperature within the operating range. |
HiF |
The temperature during the test is above the operating range. | Move to an area where the temperature is within the operating range. |
Message | What it means | What to do |
---|---|---|
Er-1 |
A used test strip was inserted. | Remove the test strip and insert a new test strip. |
Er-2 |
The blood or control solution sample was applied before the symbol appeared. | Repeat the test with a new test strip and wait until the symbol appears before applying the blood or control solution sample. |
Er-3 |
The thermometer of meter is not stabilized yet. | Retry the test after about 30 minutes for the thermometer to be stabilized. |
Er-4 |
The blood sample is of insufficient volume or blood is too thick. | Repeat the test using a new test strip. |
Er-5 |
This error message may appear when the wrong blood glucose test strip is used instead of the ReliOn™ Premier Test Strip. | Repeat the test with a ReliOn™ Premier Test Strip. Insert the test strip with the contact bars facing up, and push in gently until the meter turns on. |
Er-6 |
There is a problem with the meter. | Do not use the meter. Contact Customer Service: 1-855-776-0662. |
Er-7 |
There is a problem with the meter. This error message may also appear when the wrong blood glucose test strip is used instead of the ReliOn™ Premier Test Strip. | Repeat the test with a ReliOn™ Premier Test Strip. If the correct test strip is inserted, the meter will automatically turn on and enter test mode. If the error message persists even after inserting the correct test strip, do not use the meter and contact Customer Service: 1-855-776-0662. |
Er-8 |
There is a problem with the meter. | Do not use the meter. Contact Customer Service: 1-855-776-0662. |
Er-9 |
There is a problem with the meter. | Do not use the meter. Contact Customer Service: 1-855-776-0662. |
LoF |
The temperature during the test is below the operating range. | Move to an area where the temperature is within the operating range (50–104 ˚F) and repeat the test after the meter and test strips have reached a temperature within the operating range. |
HiF |
The temperature during the test is above the operating range. | Move to an area where the temperature is within the operating range. |